KSL-1700X-A2 is a benchtop muffle furnace up to 1700°C with fiber alumina insulation and MoSi2 heating elements.  Heating area is  200×200×200mm or 8″ x 8″ x 8″.

  • Estimated delivery time 14-30 days
  • 18 months warranty at Genuine Warranty Center.
  • Whats in the box: charging cable and block
SKU: 1700c-benchtop-muffle-furnace-988-10-l-ksl-1700x-a2

Product details

Specifications

Fsup

Furnace Structure

  • 1800 grade fiber alumina with reflective coating – max energy-saving design
  • Double-layer steel structure with air cooling fan to keep case temperature below 60°C
Heating Chamber
  • Chamber Size (W×D×H)           235×200×200 mm (9×8×8″), 9.4 L
  • Heating Space (W×D×H)          200×200×200mm  (8×8×8″), 8.4 L
  • Uniform Space ±5°C@1450°C 150×150×150mm   (6×6×6″), 3.3 L

   Power

  • 208-240VAC, 1 phase, 50/60Hz, 5.2 KW
  • 10-ft power cord without a plug

Working Temperature

  • 1650°C (continuous
  • 1700°C (< 3 hours)
  • 1800°C at heating elements

Heating Rate

  • 0~20°C / min (suggestion <= 10°C)

Temperature Accuracy

  • +/- 1°C

Heating Elements

  • Super 1800 grade MoSi2 Heating Elements (30 x 360mm)

Thermocouple

  • B Type (Pt-Rh to Pt-Rh) with Alumina tube

Temperature Controller

  • FA-YD518P-AG  PID temperature controller
  • 30 programmable temeprature segments programmed for heating, cooling and dwelling. 
  • Over-temperature alarm and thermocouple failure alarm
  • ±1°C temperature control accuracy
  • RS485 communication port. 
Eurotherm controller (optional)
  • Optional Eurotherm 3004 Temperature Controller is available upon request.
  Communication (optional)     
  • MTS01 PC communication kit is available for temperature programming, monitoring, and data logging.

Dimensions

  • W×D×H 550×540×620 mm

Net Weight

  • 75 kg (165 lbs)

Shipping Weight

  • 450 lbs

Shipping Dimensions

  • 48″ x 40″ x 49″

Warranty

  • One year limited warranty
  • Consumables, including heating elements, refractory chamber, thermocouples, are not covered by warranty. 
  • Any damages caused by corrosive and acidic materials are not covered by the warranty.
Compliance
  • CE Certified
  • NRTL or CSA certification is available upon request at an extra cost
Application Notes
  • Recommend the heating/cooling rate  ≤10°C /min.
  • Small cracks may appear on the surface of the refractory ceramics over extended use. This is a normal wear and tear
  • The refractory door block must be inserted before closing the door.
Back to Top